# Create Custom Soil Calibration

* * *

[More about soil-specific calibrations here](https://www.metergroup.com/environment/articles/how-calibrate-soil-moisture-sensors/)

* * *

To enter a soil-specific moisture calibration click the **Wrench** icon on the left.

![](https://files.helpdocs.io/la7ra8dc1e/articles/l0arv0lpvr/1592667477640/custom-calibrations-1.png)

* * *

Select **Calibrations** from the menu.

![](https://files.helpdocs.io/la7ra8dc1e/articles/l0arv0lpvr/1592667500786/custom-calibrations-2.png)

* * *

Click the **Add New Calibration** button.

![](https://files.helpdocs.io/la7ra8dc1e/articles/l0arv0lpvr/1592667515087/custom-calibrations-3.png)

* * *

Enter a name for the calibration. You can choose to **Clone** values from the existing soil calibrations and then edit them, or start with all fields blank.

![](https://files.helpdocs.io/la7ra8dc1e/articles/l0arv0lpvr/1592667534184/custom-calibrations-4.png)

* * *

Polynomial calibration example: Enter the coefficients in the corresponding sensor row and boxes for each coefficient. (Ex. a:1.17E-09, b:-3.95E-06, c:4.90E-03, d:-1.92).

![](https://files.helpdocs.io/la7ra8dc1e/articles/l0arv0lpvr/1592667566065/custom-calibrations-19.png)

Coefficients can be entered in scientific notation or full decimal form. For example 1.0e-3 or 0.001.

* * *

Linear calibration example: Enter the coefficients in the corresponding sensor row and boxes for each coefficient. (Ex. c:8.50E-04, d:-0.481).

![](https://files.helpdocs.io/la7ra8dc1e/articles/l0arv0lpvr/1592667687676/custom-calibrations-20.png)

* * *

Enter additional media parameters. Click **Accept** to save the new calibration.

![](https://files.helpdocs.io/la7ra8dc1e/articles/l0arv0lpvr/1592667717109/custom-calibrations-5.png)

The values for Bulk Density, Dry Dielectric Permittivity and Particle Density are required in order to convert **Bulk Soil Electrical Conductivity** to **Saturation Exctract** and or **Pore Water EC**.